Multiplex serum biomarker assessments: technical and biostatistical issues

Contributed equally
Journal of Translational Medicine20119:173

https://doi.org/10.1186/1479-5876-9-173

Received: 15 June 2011

Accepted: 11 October 2011

Published: 11 October 2011

Abstract

Background

Identification of predictive and prognostic biomarkers for patients with disease and undergoing different therapeutic options is a very active area of investigation. Many of these studies seek biomarkers among circulating proteins accessed in blood. Many levels of standardization in materials and procedures have been identified which can impact the resulting data.

Methods

Here, we have observed unexpected variability in levels of commonly tested analytes in serum which were processed and stored under standardized conditions. We have identified apparent changes in cytokine, chemokine and growth factor levels detected by multiplex Luminex assay in melanoma patient and healthy donor serum samples, over storage time at -80°C. Controls included Luminex kit standards, multiplexed cytokine standards and WHO cytokine controls. Data were analyzed by Wilcoxon rank-sum testing and Spearman's test for correlations.

Results

The interpretation of these changes is confounded by lot-to-lot kit standard curve reagent changes made by a single manufacturer of Luminex kits.

Conclusions

This study identifies previously unknown sources of variation in a commonly used biomarker assay, and suggests additional levels of controls needed for identification of true changes in circulating protein levels.

Keywords

serum biomarkers melanoma cytokine Luminex

Background

To improve the clinical efficacy of immunotherapies and our ability to stratify patients rationally for therapeutic intervention, biomarkers are critical to progress. The FDA's Critical Path prioritizes development of biomarkers, including a focus on aspects of: Biospecimens, Analytical Performance, Standardization and Harmonization and Bioinformatics. Accurate biomarkers offer the prospect for earlier diagnosis, improved precision of application of expensive and toxic therapies on the optimal patient populations, monitoring disease progression and therapeutic benefits as well as accelerating drug development and discovery. Guidelines for incorporation of biomarker studies in early clinical trials of novel agents have been published [1].

There is a critical need for development and validation of biomarkers to identify patients who can benefit from a particular form of immunotherapy. Only a fraction of patients benefit from IFN-α treatment [2], only a fraction of patients can achieve durable regressions in response to antigen vaccination [3], or antibody therapies, and we do not yet know the mechanisms responsible for therapeutic benefit. Despite substantial efforts from many groups, we do not know which parameters of immune response (and which assays used to assess these parameters) yield optimal results for efficacy analysis [47]. A major reason for this has been that objective clinical response rates are often below 10%, confounding the measurement of significant correlations between biomarkers and clinical responses in studies of modest size. Another important issue is that assay results may depend on biological specimen handling before assessment, and on methodological differences in complex, high throughput assays.

A number of studies in melanoma have identified candidate biomarkers of response to therapy. These range from circulating cytokines and growth factors [8, 9], gene expression profiles in tumors [10], circulating tumor cells [11], serum autoantibody profiling [12] and tumor specific T cell IFN-γ production [13] to molecular signaling pathways in tumors [14] and the nature of tumor infiltrating cells [15]. The vast majority of candidate biomarkers have not yet achieved routine clinical use due to lack of reproducibility, need for new technology and equipment, need for high quality tumor samples or high cost. The relative ease of collecting, processing, storing and shipping blood has made it a common resource for biomarker testing.

Several reports have identified phenotypic and functional changes in blood cells and serum components when the blood is held for hours or days and at different temperatures before processing [1618]. These time-dependent and temperature-dependent effects should be controlled for to the extent possible before blood processing. Standardized processing procedures by trained and competency-tested personnel can also improve immunologic assay data consistency [19]. In addition, use of freezers for sample storage that are monitored for temperature stability and that have 24 hours-a-day alarm response eliminates concerns that samples might undergo freeze-thaw cycles or be otherwise compromised by temperature changes during storage. Many of these central laboratory procedures for processing, storage and equipment maintenance are mandated by accreditation groups such as CLIA and FACT, and are described in resources from CLSI [2022].

During an investigation of biomarkers of prolonged survival after IFN-α treatment in banked melanoma patient serum samples, we discovered a number of both technical and biostatistical analysis issues [23]. Our preliminary results identified a large number of serum cytokines that appeared to correlate significantly with survival. However, further dissection of the data revealed a number of technical issues that made interpretation of the data impossible.

Here, we have performed a time course analysis of cytokines, chemokines and growth factors measured in the banked serum of healthy donors and melanoma patients stored for various intervals, and analyzed by multiplex Luminex assay. We find that a number of these analytes appear to be unstable during storage. We have also tested several aspects of the Luminex assay performance and identified a number of concerns with these multiplexed assays. Biostatistical tests indicate that despite several layers of procedural standardization and levels of controls, reliable multiplexed cytokine and chemokine determinations may be compromised by length of time in storage and/or by the changes regularly made by assay kit manufacturers to different lots and the analyte standards included. These results raise concerns about serum biomarker studies and suggest that additional controls may be required to confidently compare levels over time and between lots of reagents from the same manufacturer.

Methods

Study subjects

All serum samples were obtained after written informed consent, and under IRB approved protocols of investigation at the University of Pittsburgh. The samples received in 2005 were obtained from 23 patients at two clinical sites (Pennsylvania and Indiana). The UPCI #96-099 banking protocol was utilized for the five 2010 melanoma patient sera tested. The UPCI #04-001 healthy donor blood collection protocol was used for the blood obtained from 10 healthy donors in 2010.

Blood processing and banking

For serum collection, red top vaccutainer tubes (no anticoagulant) provided by our laboratory (Becton Dickinson #6430) in kits were used. Upon arrival in the lab, the samples are checked for proper identification, given accession numbers, and either processed immediately or (if received after 4 pm) put in the refrigerator (at 4°C) for processing the next morning. All samples were processed within 24 hours, including those drawn at external sites and shipped at ambient temperature overnight in insulated shipping containers. All processing was performed by technologists who received the same training, and the laboratory SOP #0108 was followed. Technologists also undergo annual competency training. Samples were centrifuged for 10 min at 2, 500 rpm in a refrigerated centrifuge at 4°C, then the serum was aliquoted into polypropylene freezer vials at 1.1 mL per vial and immediately placed in a -80°C freezer. All samples were stored in a monitored freezer until testing, freezer temperatures did not fluctuate above -55°C (during brief periods of high use). Samples were thawed before testing and repeated testing was performed on separate aliquots to eliminate variability from freeze-thaw cycles. The laboratory is certified under the Pennsylvania Department of Health, College of American Pathologists (CAP) and Clinical Laboratory Improvement Amendments (CLIA for Histocompatibility and General Immunology). The laboratory is registered with the FDA, and maintains a facilities master file (BB-MF-12244). The exploratory Luminex assay reported here is not used for clinical decision making, and is not a CLIA-certified assay.

Luminex assay and controls

The Luminex kits were obtained from the same manufacturer, which changed ownership during the period of the study (BioSource, Invitrogen, Life Technologies). Assays were performed only on serum samples that had been stored at -80°C. Serum samples were thawed in a refrigerator overnight (healthy donor controls, < 12 hours total time) or at room temperature the day of the assay (patient samples), clarified in a microfuge for 10 min at 1, 000 g, then diluted with the assay diluent provided per assay manufacturer's instructions. Healthy donor and control samples were run in duplicate, but large numbers of patient sera were run in singlets. The same trained technologist performed all of the assays reported herein, according to the same laboratory SOP #0037). The software used for all assays was the BioPlex System BioPlex Manager 4.0, which uses 5-parameter logistic regression. Each sample acquired ≥ 100 bead events, per manufacturers' instructions. Analytical sensitivity was calculated based on two standard deviations from the background MFI of the standard curve. There were no changes in the antibodies used for the analytes of interest reported here, and the standards were benchmarked in the same way over the time period tested here. R&D QC controls (R&D Systems QC02) are reconstituted with assay diluent from the Hu Extracellular buffer kit LHB0001 (BioSource). Each lot provides expected values for several commonly tested cytokines (as measured by R&D Systems ELISA assays). Additional kit details are presented in Additional File 1, Table S1.

To address potential inter-analysis variability, 770 data points from 2005 and 430 data points from 2010 were re-analyzed at the same time (2011) with version 6.0 software, on the original machine. There were 0/1, 200 changes in the resulting absolute values obtained.

WHO cytokine standards

WHO cytokine standards were resuspended as follows: 117187 GM-CSF WHO 88/646 10, 000 IU: contents of the ampoule were dissolved with 0.5 mL sterile distilled water and brought up to 1 mL with PBS. Further 1:10 dilution was performed with AIM V (Invitrogen) medium. 117173 IL-4 WHO 88/656 0.1 μg = 1, 000 arbitrary units per ampoule: contents of the ampule were resuspended with PBS/1% BSA, and the 1:10 dilution was performed with AIM V. 117184 IL-10 WHO 92/516 1 μg = 5, 000 RU per ampoule: contents of the ampoule were dissolved with 0.5 mL sterile distilled water and then brought up to 1 mL with PBS. Further 1:10 dilution was performed with AIM V. 117177 IL-8 WHO 89/520 1 μg = 1, 000 RU per ampoule = 1, 000, 000 pg/mL: contents of the ampoule were resuspended with PBS/1% BSA and the 1:10 dilution was performed with AIM V. To assay the WHO standards, each was diluted 1:10 (20 μL WHO standard dilution (above) + 180 μL assay diluent) and 1:50 (10 μL WHO + 490 μL assay diluent). The dilutions were treated as samples in the assay, such that the final dilutions were 1:20 and 1:100, relative to the Luminex kit standard curve (the assayed well contains 50 μL of the dilution + 50 μL of assay diluent).

Biostatistical Methods

Analyte concentrations were compared at two time points with a one-sample Wilcoxon rank-sum test on the ratio of the two concentrations. Correlation was assessed with Spearman's test. All p-values are two-sided. Assay results below the lower limit of detection or above the upper limit of quantitation were not used in the analysis.

Results and Discussion

During the analysis of a retrospective biomarker study conducted with a set of banked sera from melanoma patients [23], we discovered a potential correlation between the levels of analytes measured by Luminex and the time that the sera were stored at -80°C. Therefore, we examined several aspects of serum storage and the Luminex assay.

Repeat testing in 2010 of sera stored in 2005

Our first sample set consisted of 23 melanoma patient sera (the "old patients") who had a blood sample drawn in 2005, and had a Luminex assay performed on serum samples, on either 10/31/2005, 11/01/2005 or 2/17/2006; we refer to these as the "early" assays. To determine any changes over storage time, we thawed aliquots (not previously thawed) and tested a subset of the analytes originally tested, again by Luminex (Table 1). Unexpectedly, we identified a number of apparent changes in analyte levels. We repeated these measurements up to three times (depending on the number of previously untouched aliquots remaining) for these 23 samples: (2/02/10, 5/13/2010 and 8/11/2010)--the "late" assays. Seven of the 10 analytes we examined had highly significant changes during the approximately 5 years of storage at -80°C.
Table 1

Old patient Serum Samples

Sample

Date

Drawn

Draw

Date

Assay

Date

IL-4

pg/mL

IL-6

pg/mL

IL-8

pg/mL

IL-10

pg/mL

TNF-α

pg/mL

IFN-g

pg/mL

GM-CSF

pg/mL

IP-10

pg/mL

MIG

pg/mL

MCP-1

pg/mL

patient 1

6/2/2005

6/1/2005

10/31/2005

< 5

217

64

< 10

38

29

< 15

1339

62

10145

   

8/11/2010

25

106

353

6

< 10

28

< 15

1214

48

> 7200

   

8/11/2010

16

98

370

6

< 10

29

< 15

1195

42

> 7200

patient 2

6/2/2005

6/1/2005

10/31/2005

< 5

41

24

< 10

19

23

< 15

> 2800

130

2725

   

8/11/2010

13

20

132

7

< 10

24

< 15

> 9600

164

3149

   

8/11/2010

13

22

144

7

< 10

26

< 15

> 9600

162

2989

patient 3

6/2/2005

6/1/2005

10/31/2005

< 5

13

23

< 10

7

< 14

< 15

55

62

384

   

8/11/2010

5

5

125

5

< 10

< 5

< 15

83

90

419

   

8/11/2010

7

6

151

5

< 10

< 5

< 15

83

98

455

patient 4

7/30/2005

7/30/2005

11/1/2005

32

49

17

17

83

95

173

64

241

394

   

2/2/2010

47

17

75

20

21

63

132

87

209

633

   

5/13/2010

42

14

73

20

20

51

124

84

153

554

   

5/13/2010

42

15

69

22

21

62

133

84

154

518

   

8/11/2010

55

21

89

27

28

118

89

90

213

140

   

8/11/2010

47

23

87

33

32

144

105

90

209

128

patient 5

8/9/2005

8/9/2005

11/1/2005

12

2199

266

43

100

178

< 15

> 2800

407

> 17800

   

2/2/2010

6

1105

1469

< 14

17

159

37

2178

374

11991

   

5/13/2010

13

949

1494

< 14

16

140

36

2476

313

10275

   

5/13/2010

12

971

1428

< 14

13

121

24

2233

300

9045

patient 6

8/9/2005

8/9/2005

11/1/2005

14

592

171

28

52

229

< 15

> 2800

2586

10705

   

2/2/2010

10

350

1016

< 14

< 10

198

37

2176

2039

11703

   

5/13/2010

13

275

971

< 14

< 10

155

18

> 2980

2276

9492

   

5/13/2010

< 5

270

968

< 14

< 10

160

11

> 2980

2294

10581

patient 7

8/15/2005

8/15/2005

11/1/2005

< 5

19

47

< 5

13

< 14

< 15

285

40

2565

   

5/13/2010

8

24

242

< 14

< 10

31

< 15

445

25

5453

   

5/13/2010

< 5

20

234

< 14

< 10

27

< 15

360

25

5088

patient 8

8/20/2005

8/20/2005

11/1/2005

36

48

17

17

80

111

197

27

182

379

   

2/2/2010

46

17

99

23

23

78

146

45

172

614

   

5/13/2010

48

16

87

26

25

84

175

40

128

550

   

5/13/2010

50

15

84

22

21

70

151

41

117

544

   

8/11/2010

33

15

89

23

21

103

85

47

136

119

   

8/11/2010

49

21

107

29

30

136

95

48

150

125

patient 9

9/15/2005

9/15/2005

11/1/2005

42

67

21

11

72

122

208

275

383

1119

   

5/13/2010

77

22

78

28

36

80

224

346

328

2162

   

5/13/2010

80

22

76

27

34

87

233

347

306

2270

patient 10

9/19/2005

9/19/2005

2/17/2006

< 5

19

45

< 5

13

< 7

< 15

183

42

1656

   

5/13/2010

8

24

287

< 14

< 10

26

< 15

350

21

5509

   

5/13/2010

8

22

290

< 14

< 10

29

< 15

347

17

5021

patient 11

9/16/2005

9/22/2005

11/1/2005

48

94

2675

24

142

135

201

902

419

10026

   

5/13/2010

85

37

12663

29

39

114

272

988

300

13807

   

5/13/2010

77

37

13690

27

34

114

266

1337

285

14377

   

8/11/2010

93

41

14814

26

59

72

219

716

340

> 7200

   

8/11/2010

110

48

12823

30

58

80

238

793

364

> 7200

patient 12

9/28/2005

9/28/2005

2/17/2006

62

92

27

39

82

183

328

51

270

436

   

2/2/2010

46

22

92

36

20

118

222

88

200

909

   

5/13/2010

43

18

75

35

20

109

221

76

125

741

   

5/13/2010

34

18

88

35

22

124

208

78

165

755

patient 13

10/6/2005

10/5/2005

11/1/2005

28

67

51

8

80

77

131

877

326

6818

   

8/11/2010

83

42

342

29

55

64

225

725

332

> 7200

   

8/11/2010

116

44

335

33

72

79

228

802

356

7345

patient 14

10/7/2005

10/6/2005

11/1/2005

42

67

24

18

66

122

156

353

711

1266

   

8/11/2010

89

39

118

35

66

59

237

314

906

1088

   

8/11/2010

70

35

114

30

53

52

189

328

891

1057

patient 15

10/12/2005

---

2/17/2006

51

76

26

32

86

180

276

63

257

395

   

2/2/2010

38

21

113

48

23

146

255

107

245

801

   

5/13/2010

35

17

95

44

23

116

205

89

170

601

   

5/13/2010

28

15

100

42

20

107

224

82

168

576

patient 16

10/17/2005

---

2/17/2006

< 5

54

60

< 5

34

< 7

< 15

434

55

3950

   

5/13/2010

5

41

368

< 14

< 10

39

< 15

780

25

10159

   

5/13/2010

11

43

373

< 14

< 10

39

< 15

846

23

10552

   

8/11/2010

21

52

402

6

< 10

6

15

464

35

4616

   

8/11/2010

8

29

391

< 5

< 10

< 5

< 15

465

35

4777

patient 17

11/3/2005

11/3/2005

2/17/2006

23

52

11

< 5

49

75

111

155

283

686

   

5/13/2010

72

23

57

24

31

67

221

198

288

1734

   

5/13/2010

17

9

27

< 14

< 10

14

53

186

218

1706

   

8/11/2010

90

36

63

28

60

67

222

203

370

757

   

8/11/2010

85

30

60

24

52

56

213

202

325

705

patient 18

11/16/2005

--

2/17/2006

7

18

14

< 5

20

29

44

41

79

327

   

2/2/2010

16

15

74

< 14

< 10

34

51

59

93

807

   

5/13/2010

10

12

71

< 14

< 10

20

24

57

50

611

   

5/13/2010

8

10

72

< 14

< 10

24

30

52

48

617

patient 19

11/16/2005

---

2/17/2006

10

48

18

< 5

41

39

67

144

96

1143

   

2/2/2010

16

30

116

< 14

< 10

43

75

248

108

2812

   

5/13/2010

13

22

100

< 14

10

35

67

202

75

2101

   

5/13/2010

11

24

108

< 14

< 10

38

61

220

67

2304

patient 20

12/8/2005

12/8/2005

2/17/2006

27

31

10

< 5

62

73

149

35

162

452

   

2/2/2010

49

17

61

24

24

89

146

60

162

1422

   

5/13/2010

43

12

48

19

24

70

137

52

108

1114

   

5/13/2010

41

12

50

21

21

77

137

55

106

1234

patient 21

12/12/2005

12/9/2005

2/17/2006

< 5

24

41

< 5

16

< 7

< 15

537

35

1236

   

5/13/2010

8

22

235

< 14

< 10

29

< 15

682

25

3153

   

5/13/2010

10

23

233

< 14

< 10

27

< 15

789

25

3457

   

8/11/2010

14

29

235

11

< 10

7

< 15

518

34

1423

   

8/11/2010

13

25

238

11

< 10

6

< 15

546

39

1322

patient 22

1/26/2006

1/25/2006

2/17/2006

8

30

4

< 5

24

24

40

706

216

24

   

8/11/2010

47

22

44

17

45

37

196

332

283

888

   

8/11/2010

56

26

47

19

53

42

223

318

283

959

patient 23

1/26/2006

1/25/2006

2/17/2006

8

56

68

< 5

75

20

< 15

8705

266

75

   

8/11/2010

75

3953

534

19

56

50

202

650

339

> 7200

   

8/11/2010

76

4542

525

19

58

48

210

695

318

> 7200

     

223

        
There were different patterns seen for different groups of analytes, some of which were relatively stable over time (IL-4, change over time: p = 0.28) while others were found to change (IL-10, p = 0.093; GM-CSF, p = 0.11). Levels of some of the analytes decreased over the storage time (IL-6, p = 0.00021; decreasing in 21/23 samples; TNFα, p = 0.0078, decreasing in 20/23). Surprisingly, the IL-8 levels were significantly increased from the initial test to the subsequent tests 5 years later (IL-8, p = 0.000030, approximately 5-fold increased in 23/23 patient samples). MCP-1 levels also increased in a majority of samples (MCP-1, p = 0.00012) (Table 1/Figure 1). Each p-value was computed with a one-sample Wilcoxon test on the ratio of the 5/13/2010 assay result (for which we had the most data) to the result of the early assay.
Figure 1

Representative cytokine and chemokine changes over time. Data are shown for old patients 13 through 23, for cytokines IL-8 and IL-10, and chemokine MCP-1. On a log scale, changes detected between 2005-2006 and 2010 assays are shown.

Healthy donor and melanoma patient serum time course in 2010

To determine whether we could detect similar changes over a period of months, we drew blood from 10 healthy donors (HD, Additional File 2, Table S2, Table 2 data) and 5 melanoma patients ("new patients") (Additional File 3, Table S3, Table 3 data). HD samples were tested initially 2 months after processing and freezing, and then twice more, at 5 and 8 months of storage on the same dates as the old patient sample described above. The melanoma patient samples were tested 2 days after processing and cryopreservation, and again 3 months later.
Table 2

Healthy Donor Sera Analysis

Sample

Date

Drawn

Assay

Date

IL-4

pg/mL

IL-6

pg/mL

IL-8

pg/mL

IL-10

pg/mL

TNF-α

pg/mL

IFN-g

pg/mL

GM-CSF

pg/mL

IP-10

pg/mL

MIG

pg/mL

MCP-1

pg/mL

Healthy donor 1

12/14/2009

2/2/2010

57

43

25

21

16

44

90

28

143

324

  

2/2/2010

58

48

22

25

16

49

94

30

143

319

  

5/13/2010

50

39

19

22

13

31

87

25

103

263

  

5/13/2010

52

29

18

17

11

29

78

25

100

254

  

8/11/2010

36

42

11

25

< 10

20

40

31

71

30

  

8/11/2010

29

39

11

20

< 10

16

39

29

64

31

Healthy donor 2

12/16/2009

2/2/2010

57

61

44

41

20

160

179

31

96

632

  

2/2/2010

59

59

40

42

18

163

151

31

105

589

  

5/13/2010

44

48

29

31

15

109

129

28

70

529

  

5/13/2010

41

49

34

33

15

109

128

27

69

521

  

8/11/2010

82

80

63

37

42

100

84

40

151

160

  

8/11/2010

67

91

68

39

34

105

81

35

165

138

Healthy donor 3

12/17/2009

2/2/2010

17

< 8

18

< 14

< 10

< 12

< 15

23

20

977

  

2/2/2010

21

< 8

19

< 14

< 10

< 12

< 15

23

13

921

  

5/13/2010

22

< 8

21

< 14

< 10

< 12

< 15

22

< 12

803

  

5/13/2010

20

< 8

18

< 14

< 10

< 12

< 15

23

< 12

763

  

8/11/2010

24

< 3

18

< 5

< 10

< 5

< 15

19

11

241

  

8/11/2010

32

< 3

21

< 5

< 10

< 5

< 15

23

16

258

Healthy donor 4

12/18/2009

2/2/2010

111

29

88

51

50

189

253

39

196

577

  

2/2/2010

121

31

90

51

56

212

262

37

216

579

  

5/13/2010

81

20

66

38

39

147

211

30

128

468

  

5/13/2010

76

21

60

36

33

142

201

30

133

440

  

8/11/2010

232

48

160

65

90

137

173

49

277

171

  

8/11/2010

222

46

167

68

92

141

167

48

276

183

Healthy donor 5

12/21/2009

2/2/2010

12

9

33

< 14

< 10

18

33

20

32

194

  

2/2/2010

< 5

< 8

30

< 14

< 10

14

25

20

13

192

  

5/13/2010

5

< 8

34

< 14

< 10

14

< 15

19

< 12

177

  

5/13/2010

< 5

< 8

32

< 14

< 10

< 12

< 15

20

12

174

  

8/11/2010

7

7

26

8

< 10

> 5

< 15

25

20

16

  

8/11/2010

12

12

33

10

< 10

< 5

23

27

20

17

Healthy donor 6

12/21/2009

2/2/2010

19

8

24

< 14

< 10

< 12

37

22

50

496

  

2/2/2010

< 5

< 8

28

< 14

< 10

< 12

17

22

37

558

  

5/13/2010

19

< 8

15

< 14

< 10

< 12

18

19

40

434

  

5/13/2010

10

< 8

< 12

< 14

< 10

< 12

18

18

36

413

  

8/11/2010

12

12

47

12

14

7

21

27

48

111

  

8/11/2010

18

11

45

13

13

7

21

28

51

108

Healthy donor 7

12/22/2009

2/2/2010

16

< 8

38

< 14

< 10

35

56

19

66

1040

  

2/2/2010

17

9

40

< 14

10

35

62

19

66

1019

  

5/13/2010

19

< 8

35

< 14

< 10

40

51

17

50

843

  

5/13/2010

20

< 8

41

< 14

< 10

33

51

18

53

848

  

8/11/2010

13

6

28

28

< 10

19

37

21

24

245

  

8/11/2010

16

10

30

41

11

25

49

24

24

246

Healthy donor 8

12/23/2009

2/2/2010

54

15

39

29

17

82

135

40

188

926

  

2/2/2010

58

17

33

29

17

78

123

42

188

934

  

5/13/2010

64

15

37

32

21

76

144

38

160

815

  

5/13/2010

65

15

35

33

20

72

129

36

160

742

  

8/11/2010

23

< 3

10

8

< 10

9

22

34

71

144

  

8/11/2010

32

6

14

14

< 10

21

36

34

96

130

Healthy donor 9

12/24/2009

2/2/2010

< 5

9

17

< 14

< 10

< 12

< 15

21

13

969

  

2/2/2010

< 5

8

15

< 14

< 10

< 12

< 15

20

20

928

  

5/13/2010

< 5

< 8

13

< 14

< 10

< 12

< 15

17

< 12

784

  

5/13/2010

< 5

< 8

14

< 14

< 10

< 12

< 15

19

< 12

813

  

8/11/2010

8

11

19

9

< 10

6

< 15

29

20

332

  

8/11/2010

7

10

17

8

< 10

< 5

< 15

26

20

331

Healthy donor 10

12/28/2009

2/2/2010

< 5

8

< 12

< 14

< 10

< 12

< 15

37

13

1034

  

2/2/2010

< 5

< 8

< 12

< 14

< 10

16

< 15

37

13

990

  

5/13/2010

< 5

< 8

< 12

< 14

< 10

< 12

< 15

34

< 12

845

  

5/13/2010

< 5

< 8

< 12

< 14

< 10

< 12

< 15

36

< 12

802

  

8/11/2010

< 5

4

< 3

< 5

< 10

< 5

< 15

57

6

374

  

8/11/2010

5

5

< 3

< 5

< 10

8

< 15

59

11

385

Table 3

New Melanoma Patient Sera Analysis

Sample

Draw

Date

Assay

Date

IL-4

pg/mL

IL-6

pg/mL

IL-8

pg/mL

IL-10

pg/mL

TNF-α

pg/mL

IFN-g

pg/mL

GM-CSF

pg/mL

IP-10

pg/mL

MIG

pg/mL

MCP-1

pg/mL

Mel. Pt. 1

5/10/2010

5/13/2010

15

11

42

< 14

95

24

< 15

39

12

754

 

5/10/2010

5/13/2010

10

8

39

< 14

82

22

< 15

40

17

754

 

5/10/2010

8/11/2010

21

12

44

22

183

20

41

49

32

270

 

5/10/2010

8/11/2010

24

13

43

29

168

23

51

50

32

268

Mel. Pt. 2

5/10/2010

5/13/2010

18

17

87

< 14

< 10

26

< 15

30

21

1437

 

5/10/2010

5/13/2010

13

16

95

< 14

< 10

21

< 15

31

17

1494

 

5/10/2010

8/11/2010

28

30

97

22

10

25

33

38

39

664

 

5/10/2010

8/11/2010

25

31

86

24

10

24

30

37

28

662

Mel. Pt. 3

5/10/2010

5/13/2010

42

21

72

29

20

< 12

18

190

81

771

 

5/10/2010

5/13/2010

38

19

70

25

17

< 12

18

188

78

732

 

5/10/2010

8/11/2010

34

14

73

31

38

9

47

141

96

223

 

5/10/2010

8/11/2010

36

15

70

28

32

6

50

135

89

198

Mel. Pt. 4

5/10/2010

5/13/2010

68

26

45

51

19

62

107

19

145

955

 

5/10/2010

5/13/2010

66

24

42

53

18

63

111

18

143

875

 

5/10/2010

8/11/2010

102

36

26

120

< 10

16

50

19

100

238

 

5/10/2010

8/11/2010

99

41

24

139

< 10

16

48

20

103

215

Mel. Pt. 5

5/10/2010

5/13/2010

35

64

380

< 14

13

27

78

27

106

831

 

5/10/2010

5/13/2010

34

61

393

< 14

10

27

82

26

101

737

 

5/10/2010

8/11/2010

33

43

458

20

16

27

54

32

139

170

 

5/10/2010

8/11/2010

45

53

480

25

24

34

55

38

146

222

As expected, HD samples had low circulating levels of many analytes tested. These HD control samples also showed changes in analyte levels, even after short-term storage. Again, some analytes were stable, others were much less stable. IL-8 increased in 3/10 HD, at the 8 month timepoint (n.s.), but not by 5 months. IP-10 also began to increase in 5/10 HD at 8 months (p = 0.01). Several analytes decreased in the relatively short storage time interval, including IFNγ (p = 0.06 at 5 mo., p = 0.03 at 8 mo., decreasing in 6/10 HD), and MCP-1, which showed the most dramatic decreases in 10/10 donors, by 8 mo. (p = 0.002). These changes, between the first assay and the second and third assays (100 and 190 days apart), are shown graphically in Figure 2. The melanoma patient samples did not show significant changes within the short storage time, with the exception of MCP-1, which decreased in 5/5 samples within 3 months (p = 0.06). When the ratios of the concentrations of the different analytes measured at different times were plotted together (Figure 3), the trends in concentration changes observed were not significantly different between the serum sample data sets (old patients, HD, new patients) (Table 1, Table 2, Table 3).
Figure 2

Time course of analyte concentrations for healthy donors. Assays done on 5/13/2010 and 8/11/2010 were normalized to those done on 2/02/2010.

Figure 3

Comparison of assay results obtained with sera from healthy donors, new melanoma patients and old melanoma patients. Points are the ratio of concentrations of the assays done on 8/11/2010 normalized to those done on 5/13/2010.

Cytokine Controls used in assays

We purchased our Luminex kits from a single source, however, that source changed ownership between Oct. '05 and Aug. '10 (from Biosource to Invitrogen to Life Technologies). Each kit includes reagents to generate an 8-point standard curve from which all values are determined. For the custom kits we requested, to test a specific array of analytes of interest, the manufacturer pre-tests the specific antibodies together, to confirm lack of cross-reactivity. The manufacturer indicates that the kits are not released unless the following criteria are met: " < 10% cross-reactivity to related recombinant protein at the highest point of the standard curve" (Life Technologies). We requested the specific cross-reactivity testing data performed for the kits we used in this study, but were repeatedly informed that company policy prohibits QC data release to customers.

As an additional control, we included "Multiplex QC" controls, which are complex mixtures of recombinant cytokines, chemokines and growth factors prepared by the manufacturer at 3 concentrations (low, medium and high). We have established the reproducibility of this control (Additional File 4, Table S4) when tested via Luminex (% CV = 1%-52%, average % CV = 14% for 8 analytes). While the absolute values for each analyte do not exactly match the "expected" value from the QC control manufacturer (R&D Systems), they are similar, and we use a different platform and different antibody clones for detection via Luminex, which may account for those differences (as indicated in the package insert).

We also received WHO cytokine standards for IL-4, IL-8, IL-10 and GM-CSF. These lyophilized cytokine controls were resuspended (Materials and Methods) and individually tested at 1:10, 1:50 and 1:100 dilutions in two replicate Luminex assays for the same ten analytes described above. These data are presented in Table 4. As expected, the standard under study was almost always detected. However, there were some surprising results. MCP-1 was also almost always detected in addition to the standard, and MIG was always detected when the standard IL-10 was used. The apparent concentrations of these two analytes in some instances exceeded 10% of that of the standard. IL-6, IFN-γ and GM-CSF also showed evidence of minor cross-reactivity.
Table 4

WHO Cytokine Standards

Lab Number

Assay

Date

IL-4

pg/mL

IL-6

pg/mL

IL-8

pg/mL

IL-10

pg/mL

TNF-α

pg/mL

IFN-g

pg/mL

GM-CSF

pg/mL

IP-10

pg/mL

MIG

pg/mL

MCP-1

pg/mL

117173 IL-4

           

1:10

40311

17497

N/A

N/A

N/A

N/A

N/A

N/A

N/A

N/A

N/A

1:10

40401

11364

N/A

N/A

N/A

N/A

N/A

N/A

N/A

N/A

196

1:10

40401

10956

N/A

N/A

N/A

N/A

N/A

N/A

N/A

N/A

114

1:50

40311

10945

N/A

N/A

N/A

N/A

N/A

N/A

N/A

N/A

N/A

1:50

40401

1350

N/A

N/A

N/A

N/A

N/A

N/A

N/A

N/A

392

1:50

40401

1321

N/A

N/A

N/A

N/A

N/A

N/A

N/A

N/A

1220

117177 IL-8

           

1:10

40311

N/A

N/A

216983

N/A

N/A

N/A

N/A

N/A

N/A

N/A

1:10

40401

N/A

N/A

153880

N/A

N/A

N/A

N/A

N/A

N/A

563

1:10

40401

N/A

N/A

153707

N/A

N/A

N/A

N/A

N/A

N/A

509

1:50

40311

N/A

N/A

QA

N/A

N/A

N/A

N/A

N/A

N/A

N/A

1:50

40401

N/A

N/A

45621

N/A

N/A

N/A

N/A

N/A

N/A

2169

1:50

40401

N/A

N/A

46708

N/A

N/A

N/A

N/A

N/A

N/A

1445

117184 IL-10

           

1:10

40311

N/A

N/A

N/A

119338

N/A

N/A

180

N/A

1813

N/A

1:10

40401

N/A

230

N/A

72096

N/A

N/A

N/A

N/A

3621

318

1:10

40401

N/A

226

N/A

95800

N/A

N/A

N/A

N/A

3891

389

1:50

40311

N/A

N/A

N/A

95462

N/A

N/A

N/A

N/A

3836

N/A

1:50

40401

N/A

340

N/A

39419

N/A

N/A

N/A

N/A

4488

1855

1:50

40401

N/A

179

N/A

30223

N/A

N/A

N/A

N/A

4053

1308

117187GM-CSF

           

1:10

40401

N/A

N/A

N/A

N/A

N/A

373

75824

N/A

N/A

875

1:10

40401

N/A

N/A

N/A

N/A

N/A

272

70453

N/A

N/A

721

1:10

40311

N/A

N/A

N/A

N/A

N/A

N/A

47332

N/A

N/A

N/A

1:10

40311

N/A

N/A

N/A

N/A

N/A

N/A

78882

N/A

N/A

769

1:50

40311

N/A

N/A

N/A

N/A

N/A

N/A

59603

N/A

N/A

N/A

1:50

40311

N/A

N/A

N/A

N/A

N/A

N/A

76623

N/A

N/A

1063

1:50

40311

N/A

N/A

N/A

N/A

N/A

N/A

19209

N/A

N/A

1193

The apparent cross-reactivity seen for MCP-1 and MIG might be caused by a medium additive present in the AIM V medium (a serum-free lymphocyte culture medium) used in a dilution step for these proteins. We tested several commonly used culture medias (AIM V, RPM1640, Iscoves and CellGenix DC media) in a 30-plex Luminex assay which also included a repeat test of the WHO standards. The results did identify low levels (3-62 pg/mL) of several analytes in the culture medias (HGF, FGF basic, RANTES, IL-17 and IL2R) but not MCP-1 or MIG (data not shown). The MCP-1 was again detected in the IL-8 and GM-CSF WHO standards and MIG in the IL-10 standard (as well as HGF, FGF basic and RANTES). We are investigating other possible sources of low levels of other cytokines and growth factors in the WHO standards.

As a test of the day-to-day reproducibility of two of the cytokines of particular interest, IL-6 and IL-8, a set of samples and controls were run in two different custom kits one day apart (with samples kept thawed, at 4°C overnight), in which both IL-6 and IL-8 were included in both kits. Notably, these two kits also had different standard curves and upper limits of detection. For IL-6, the 10-plex kit upper limit was 7, 400 pg/mL, while in the 8-plex, it was 13, 800 pg/mL (1.8 fold higher). For IL-8, the 10-plex upper limit was 24, 800 pg/mL and in the 8-plex, 10, 160 pg/mL (2.4 fold lower). When the values for the 38 samples were compared between the two kits, the ratio of the IL-6 values was 1.0 (median & mean), showing excellent concordance. For IL-8, where the upper limits were more disparate, the ratio of the values was 0.80, which was a small but significant difference (Figures 4A and 4B). These data indicate that the assay with the higher upper limit has larger measured values.
Figure 4

Two plates run together compared for A) IL-6 and B) IL-8 values. A set of 38 cell culture samples were run on both an 8-plex and a 10-plex plate. The values for IL-6 and IL-8 are compared on a log scale. Each plate had a unique upper limit. The values for IL-6 show excellent concordance, and the 8-plex upper limit was 1.8 times the upper limit. The IL-8 values were reproducibly higher (1.25×) in the 10-plex plate where the upper limit was 2.4 times higher.

Upper limit problem

The Luminex kits that we used at the different time points were not identical. In particular, we noticed that the upper limits of quantitation for individual analytes changed over time for the different kits. In principal, this should not affect the measured concentrations, because the kits include kit-specific standards to generate 8-point standard curves matched to the expected detection range. However, if the concentration determinations were affected, that would confound our interpretation of the observed changes in analyte concentration over time, and therefore we investigated that possibility. Data from assays done on 5/13/2010 ("late" assay) were compared to data from assays on 10/31/2005, 11/1/2005 or 2/17/2006 ("early" assays). Kits used in 2005 and 2006 had the same upper limits, and because no samples had assays done on the same date, results were combined. Figure 5 is a scatter plot of the late-to-early ratio of analyte concentrations versus the late-to-early ratio of assay upper limits assays with a smooth curve is superimposed. The late-to-early ratio of upper limits was different for each of the 10 analytes. Typically, 12 samples were assessed for each analyte. The correlation of the two ratios is highly significant (p < 10-15, Spearman's test). Therefore, we are concerned that assays performed at different times with different kits may not be comparable.
Figure 5

Scatter plot of the late-to-early ratio of analyte concentrations versus the late-to-early ratio of assay upper limits of quantitation; a smooth curve is superimposed. Early assays were done on 10/31/2005, 11/01/2005 or 2/17/2006; late assays were done on 5/13/2010.

In this report, we detail reproducibility problems we encountered testing circulating cytokines, chemokines and growth factors by Luminex in serum samples which were stored over months to years under highly controlled conditions. Some of these changes were very dramatic: IL-8 increased 4-6 fold in old patient samples; MCP-1 decreased 4-6 fold in new patient samples, and up to 10-fold in healthy donor samples; IL-10 changed from negative to positive or positive to negative within the same old patient serum dataset (Figure 1). Our initial hypothesis was that the changes were entirely biological, and that despite standardized blood handling procedures and temperature-controlled freezer storage, some analytes became unstable over time or upon thaw. Two recent reports testing cytokine stability found most tested cytokines to be stable over 1-2 years at -80°C, and a subset (including IL-8 and IL-10) became unstable after 2-4 years [24, 25]. Many of the proteins became unstable after repeated freeze-thaw cycles. If these were the only mechanisms, then the analytes we tested should have behaved consistently between our three datasets, because the change would be analyte-specific. This is not the only explanation, because, for example, MCP-1 increased over time in the majority of old patient samples and decreased over time in both HD and new patient sets.

Our study has a number of limitations. The more recently acquired HD and new patient data sets were tested within months of blood draw. A better analysis of the impact of storage time on analyte stability would require a large number of patients and HD samples stored for longer periods with costly repeated multiplex testing. We also limited the diversity of analytes we examined. Another variable was the time from blood draw to serum separation and freezing. Some of our samples were drawn within the laboratory and at our nearby clinic and processed within a few hours, while other old patient samples were shipped overnight and processed the following morning. However, the nature of these blood handling procedures reflects the unavoidable limitations inherent in transferring patient blood from the clinic to a central laboratory capable of standardized processing, as well as for multi-institutional trials where large numbers of patients can be treated and tested, but overnight shipping is required. Lastly, some of our healthy donor and control samples were run in duplicate, but to reduce costs, large numbers of patient sera were run in singlets. Due to the small average % CVs determined for many duplicates (Additional File 1, Table S1) this may have minimal impact on the trends we observed.

The Luminex assay has been shown (by ourselves [26] and others [27]) to show good correspondence to ELISA platform assays. In addition, the Luminex assay has good reproducibility from well-to-well, and from day-to-day (Figure 4). Also, our use of the R&D QC controls (Additional File 4, Table S4) indicate good reproducibility of recombinant analytes when mixed together. This may indicate that the serum matrix may impact reproducibility, and/or the biological impact of a tumor may lead to systemic changes (including altered glycosylation) which impact the assay.

This study also suggests that the changes in the upper limits of detection, which can vary substantially from kit to kit, month to month, and analyte to analyte from a single manufacturer, may impact the ability to determine analyte concentration. This impacts kit-to-kit reproducibility, and greatly increases the importance of comparing samples with the identical lot of kits with identical standard curve ranges. We attempted to dissect this further by requesting access to manufacturer QC data, but we were repeatedly denied access to any additional information specific to the testing performed on the kits we used.

We do not understand why the assay kit upper limits seem to affect assay performance in the systematic way that is evident in Figure 5. However, we have to conclude that the results of assays done with different kits cannot be directly compared. Therefore, the apparent changes in analyte levels over time that we observe may arise from the kit-to-kit variability: we cannot claim to observe changes in analyte levels over storage time at -80°C.

Conclusions

In conclusion, the multiplex Luminex platform offers the opportunity to test a wide variety of analytes in the same sample, with minimal volume requirements, and good well-to-well and day-to-day reproducibility. These attributes are important when broadly searching for serum biomarkers. However, we find that a number of commonly tested candidate immunologic biomarkers show evidence of unexpected, large variability when tested retrospectively, after long storage times. This variability can be reduced by 1) performing assays with kits from a single lot, and potentially 2) minimizing storage time before retrospective analysis of banked serum.

Notes

List of Abbreviations

IL: 

interleukin

TNF: 

tumor necrosis factor

GM-CSF: 

granulocyte-macrophage colony stimulating factor

PBMC: 

peripheral blood mononuclear cells

FDA: 

Food and Drug Administration

FBS: 

fetal bovine serum.

Declarations

Acknowledgements and funding

We acknowledge Sharon Sember (IML) for conduct of the Luminex assays (UPCI IML), and Dr. Theresa L. Whiteside (University of Pittsburgh) for helpful discussions.

This study was supported in part by the University of Pittsburgh Cancer Institute and the NIH Cancer Center Support Grant P30 CA047904; NCI RO1 CA138635 (LHB); Developmental Research Funds of the SPORE in Skin Cancer P50 CA121973 (JMK); Frontier Science and Technology Research Foundation and ECOG Central Laboratory Support (LHB).

Authors’ Affiliations

(1)
University of Pittsburgh Cancer Institute
(2)
University of Pittsburgh School of Medicine
(3)
Department of Medicine, University of Pittsburgh
(4)
Department of Surgery, University of Pittsburgh
(5)
Department of Immunology, University of Pittsburgh
(6)
Depament of Biostatistics, University of Pittsburgh

References

  1. Dancey JE, Dobbin KK, Groshen S, Jessup JM, Hruszkewycz AH, Koehler M, Parchment R, Ratain MJ, Shankar LK, Stadler WM, True LD, Gravell A, Grever MR, Biomarkers Task Force of the NCI Investigational Drug Steering Committee: Guidelines for the development and incorporation of biomarker studies in early clinical trials of novel agents. Clin Cancer Res. 2010, 16: 1745-55. 10.1158/1078-0432.CCR-09-2167.View ArticlePubMedGoogle Scholar
  2. Kirkwood JM, Manola J, Ibrahim J, Sondak V, Ernstoff MS, Rao U, Eastern Cooperative Oncology Group: A pooled analysis of eastern cooperative oncology group and intergroup trials of adjuvant high-dose interferon for melanoma. Clin Cancer Res. 2004, 10: 1670-1677. 10.1158/1078-0432.CCR-1103-3.View ArticlePubMedGoogle Scholar
  3. Rosenberg SA, Restifo NP, Yang JC, Morgan RA, Dudley ME: Adoptive cell transfer: a clinical path to effective cancer immunotherapy. Nat Rev Cancer. 2008, 8: 299-308. 10.1038/nrc2355.PubMed CentralView ArticlePubMedGoogle Scholar
  4. Butterfield LH, Disis ML, Fox BA, Lee PP, Khleif SN, Thurin M, Trinchieri G, Wang E, Wigginton J, Chaussabel D, Coukos G, Dhodapkar M, Hakansson L, Janetzki S, Kleen TO, Kirkwood JM, Maccalli C, Maecker H, Maio M, Malyguine A, Masucci G, Palucka AK, Potter DM, Ribas A, Rivoltini L, Schendel D, Seliger B, Selvan S, Slingluff CL, Stroncek DF, Streicher H, Wu X, Zeskind B, Zhao Y, Zocca M-B, Zwierzina H, Marincola FM: A systematic approach to biomarker discovery; Preamble to "the iSBTc-FDA taskforce on immunotherapy biomarkers. J Transl Med. 2008, 6: 81-10.1186/1479-5876-6-81.PubMed CentralView ArticlePubMedGoogle Scholar
  5. Tahara H, Sato M, Thurin M, Wang E, Butterfield LH, Disis ML, Fox BA, Lee PP, Khleif SN, Wigginton JM, Ambs S, Akutsu Y, Chaussabel D, Doki Y, Eremin O, Fridman WH, Hirohashi Y, Imai K, Jacobson J, Jinushil M, Kanamoto A, Kashani-Sabet M, Kato K, Kawakami Y, Kirkwood JM, Kleen TO, Lehmann PV, Liotta L, Lotze MT, Malyguine A, Masucci G, Matsubara H, Nakamura K, Nishikawa H, Palucka AK, Petricoin EF, Pos Z, Ribas A, Rivoltini L, Sato N, Shiku H, Slingluff CL, Streicher H, Stronchek DF, Takeuchi H, Toyota M, Wada H, Wu X, Wulfkuhle J, Yaguchi T, Zeskind B, Zhao Y, Zocca M-B, Marincola FM: Emerging concepts in biomarker discovery: The US-Japan workshop on immunological molecular markers in oncology. J Transl Med. 2009, 7: 45-10.1186/1479-5876-7-45.PubMed CentralView ArticlePubMedGoogle Scholar
  6. Butterfield LH, Disis ML, Khleif SN, Balwit JM, Marincola F: Immuno-oncology biomarkers 2010 and beyond: Perspectives from the iSBTc/SITC Biomarker Task Force. J Transl Med. 2010, 8: 130-10.1186/1479-5876-8-130.PubMed CentralView ArticlePubMedGoogle Scholar
  7. Butterfield LH, Palucka AK, Britten CM, Dhodapkar MV, Hakansson L, Janetzki S, Kawakami Y, Kleen T-O, Lee PP, Macalli C, Maecker HT, Maino VC, Maio M, Malyguine A, Masucci G, Pawelec G, Potter DM, Rivoltini L, Salazar LG, Schendel DJ, Slingluff CL, Song W, Stroncek DF, Tahara H, Thurin M, Trinchieri G, van Der Burg SH, Whiteside TL, Wigginton JM, Marincola F, Khleif S, Fox BA, Disis ML: Recommendations from the iSBTc-SITC/FDA/NCI Workshop on Immunotherapy Biomarkers. Clin Cancer Res. 2011, 17: 3064-3076. 10.1158/1078-0432.CCR-10-2234.PubMed CentralView ArticlePubMedGoogle Scholar
  8. Yurkovetsky ZR, Kirkwood JM, Edington HD, Marrangoni AM, Velikokhatnaya L, Winans MT, Gorelik E, Lokshin AE: Multiplex analysis of serum cytokines in melanoma patients treated with interferon-alpha2b. Clin Cancer Res. 2007, 13: 2422-2428. 10.1158/1078-0432.CCR-06-1805.View ArticlePubMedGoogle Scholar
  9. Hutchinson PE, Osborne JE, Pringle JH: Higher serum 25-hydroxy vitamin D3 levels at presentation are associated with improved survival from melanoma, but there is no evidence that later prevailing levels are protective. J Clin Oncol. 2010, 28: e492-493. 10.1200/JCO.2010.29.6095.View ArticlePubMedGoogle Scholar
  10. Gajewski TF, Fuertes M, Spaapen R, Zheng Y, Kline J: Molecular profiling to identify relevant immune resistance mechanisms in the tumor microenvironment. Curr Opin Immunol. 2011, 23: 286-292. 10.1016/j.coi.2010.11.013.PubMed CentralView ArticlePubMedGoogle Scholar
  11. Ma J, Lin JY, Alloo A, Wilson BJ, Schatton T, Zhan Q, Murphy GF, Waaga-Gasser AM, Gasser M, Stephen Hodi F, Frank NY, Frank MH: Isolation of tumorigenic circulating melanoma cells. Biochem Biophys Res Commun. 2010, 402: 711-717. 10.1016/j.bbrc.2010.10.091.PubMed CentralView ArticlePubMedGoogle Scholar
  12. Liu Y, He J, Xie X, Su G, Teitz-Tennenbaum S, Sabel MS, Lubman DM: Serum autoantibody profiling using a natural glycoprotein microarray for the prognosis of early melanoma. J Proteome Res. 2010, 9: 6044-6051. 10.1021/pr100856k.PubMed CentralView ArticlePubMedGoogle Scholar
  13. Kirkwood JM, Lee S, Moschos SJ, Albertini MR, Michalak JC, Sander C, whiteside TL, Butterfield LH, Weiner L: Immunogenicity and antitumor effects of vaccination with peptide vaccine+/-granulocyte-monocyte colony-stimulating factor and/or IFN-alpha2b in advanced metastatic melanoma: Eastern Cooperative Oncology Group Phase II Trial E1696. Clin Cancer Res. 2009, 15: 1443-1451. 10.1158/1078-0432.CCR-08-1231.PubMed CentralView ArticlePubMedGoogle Scholar
  14. Gould Rothberg BE, Rimm DL: Biomarkers: the useful and the not so useful--an assessment of molecular prognostic markers for cutaneous melanoma. J Invest Dermatol. 2010, 130: 1971-1987. 10.1038/jid.2010.149.View ArticlePubMedGoogle Scholar
  15. Hodi FS, Dranoff G: The biologic importance of tumor-infiltrating lymphocytes. J Cutan Pathol. 2010, 37 (Suppl 1): 48-53.PubMed CentralView ArticlePubMedGoogle Scholar
  16. Bull M, Lee D, Stucky J, Chiu YL, Rubin A, Horton H, McElrath MJ: Defining blood processing parameters for optimal detection of cryopreserved antigen-specific responses for HIV vaccine trials. J Immunol Meth. 2007, 322: 57-69. 10.1016/j.jim.2007.02.003.View ArticleGoogle Scholar
  17. Kierstead LS, Dubey S, Meyer B, Tobery TW, Mogg R, Fernandez VR, Long R, Guan L, Gaunt C, Collins K, Sykes KJ, Mehrotra DV, Chirmule N, Shiver JW, Casimiro DR: Enhanced rates and magnitude of immune responses detected against an HIV vaccine: effect of using an optimized process for isolating PBMC. AIDS Res Hum Retroviruses. 2007, 23: 86-92. 10.1089/aid.2006.0129.View ArticlePubMedGoogle Scholar
  18. McKenna KC, Beatty KM, Bilonick RA, Schoenfield L, Lathrop KL, Singh AD: Activated CD11b+CD15+ granulocytes increase in the blood of patients with uveal melanoma. Invest Ophthalmol Vis Sci. 2009, 50: 4295-4303. 10.1167/iovs.08-3012.View ArticlePubMedGoogle Scholar
  19. Boaz MJ, Hayes P, Tarragona T, Seamons L, Cooper A, Birungi J, Kitandwe P, Semaganda A, Kaleebu P, Stevens G, Anzala O, Farah B, Ogola S, Indangasi J, Mhlanga P, Van Eeden M, Thakar M, Pujari A, Mishra S, Goonetilleke N, Moore S, Mahmoud A, Sathyamoorthy P, Mahalingam J, Narayanan PR, Ramanathan VD, Cox JH, Dally L, Gill DK, Gilmour J: Concordant proficiency in measurement of T-cell immunity in human immunodeficiency virus vaccine clinical trials by peripheral blood mononuclear cell and enzyme-linked immunospot assays in laboratories from three continents. Clin Vacc Immunol. 2009, 16: 147-155. 10.1128/CVI.00326-08.View ArticleGoogle Scholar
  20. Clinical Laboratory Improvement Amendments (CLIA) Brochures. [http://www.cms.hhs.gov/CLIA/]
  21. International Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human Use (ICH). [http://www.ich.org]
  22. FDA Food and Drug Administration: Draft Guidance for Industry: Potency Tests for Cellular and Gene Therapy Products. [http://www.fda.gov/BiologicsBloodVaccines/GuidanceComplianceRegulatoryInformation/Guidances/CellularandGeneTherapy/ucm072571.htm]
  23. Potter DM, Butterfield LH, Divito SJ, Sander C, Kirkwood JM: Pitfalls in retrospective analyses of biomarkers: a case study with metastatic melanoma patients. 2011,Google Scholar
  24. de Jager W, Bourcier K, Rijkers GT, Prakken BJ, Seyfert-Margolis V: Prerequisites for cytokine measurements in clinical trials with multiplex immunoassays. BMC Immunol. 2009, 10: 52-10.1186/1471-2172-10-52.PubMed CentralView ArticlePubMedGoogle Scholar
  25. Zhou XB, Fragala MS, McElhaney JEB, Kuchel GA: Conceptual and methodological issues relevant to cytokine and inflammatory marker measurements in clinical research. Curr Opin Clin Nutr Metabolic Care. 2010, 13: 541-547. 10.1097/MCO.0b013e32833cf3bc.View ArticleGoogle Scholar
  26. Butterfield LH, Gooding W, Whiteside TL: Development of a potency assay for human dendritic cells: IL-12p70 production. J Immunother. 2008, 31: 89-100. 10.1097/CJI.0b013e318158fce0.View ArticlePubMedGoogle Scholar
  27. Khan SS, Smith MS, Reda D, Suffredini AF, McCoy JP: Multiplex bead array assays for detection of soluble cytokines: comparisons of sensitivity and quantitative values among kits from multiple manufacturers. Cytometry B Clin Cytom. 2004, 61: 35-39.View ArticlePubMedGoogle Scholar

Copyright

© Butterfield et al; licensee BioMed Central Ltd. 2011

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.