Table 5 Details of instrument output discrepancies with statistical significance

From: Bladder cancer risk stratification with the Oncuria 10-plex bead-based urinalysis assay using three different Luminex xMAP instrumentation platforms

Biomarker

Instruments

Mean 1 (pg/mL)

Mean 2 (pg/mL)

∆ (pg/mL)

% Differencea (%)

95% CI (pg/mL)

P-value, adjusted†

ApoE

200 vs FM3D

819.5

796.1

22.4

2.8

4.6 to 40.1

0.0107

MMP-9

MPX vs 200

166.7

175.1

− 8.4

5.0

− 14.0 to − 2.9

0.0020

 

MPX vs FM3D

166.7

174.1

− 7.4

4.4

− 14.0 to − 0.8

0.0261

SDC1

MPX vs 200

13,986.0

13,690.0

295.3

2.2

91.9 to 498.8

0.0031

 

200 vs FM3D

13,690.0

13,991.0

− 301.1

2.2

− 494.2 to − 108.1

0.0015

  1. aDifferences calculated as the absolute value of the ∆ divided by the lower of the two mean protein values × 100, with % differences rounded to nearest 0.1%
  2. †P-values calculated by repeated measures ANOVA with Tukey post-test correction for multiple comparisons. All concentration values rounded to nearest 0.1 picogram
  3. Instrument Abbreviations: 200 = Luminex 200; FM3D = FlexMap 3D; MPX = MagPix