Skip to main content

Table 4 Biomarker protein concentrations in urine samples compared across three analyzers (pg/mL)

From: Bladder cancer risk stratification with the Oncuria 10-plex bead-based urinalysis assay using three different Luminex xMAP instrumentation platforms

Sample ID

Instrument

A1AT

ANG

ApoE

CA9

IL-8

MMP-9

MMP-10

PAI-1

SDC1

VEGF-A

#0003

MagPix

185,250

4320

10,566

49

1410

1610

709

4356

38,128

6225

Tumor

200

185,250

4320

11,668

44

1410

1652

720

4639

38,657

6545

 

FlexMap 3D

185,250

4320

11,707

41

1410

1694

761

4649

40,278

6646

#0146

MagPix

185,250

1952

2618

11

855

161

13

1889

29,133

1496

Tumor

200

185,250

1922

2823

6

893

179

13

1832

27,662

1535

 

FlexMap 3D

185,250

1978

2728

10

903

170

13

1914

29,256

1567

#0147

MagPix

185,250

479

204

1

138

36

13

103

11,336

50

Tumor

200

185,250

475

204

1

140

46

13

109

11,115

57

 

FlexMap 3D

185,250

464

204

1

139

40

13

106

11,111

53

#0010

MagPix

3570

28

204

1

2

21

13

8

3099

25

Control

200

3872

25

204

4

2

21

13

8

3306

29

 

FlexMap 3D

3581

28

204

1

2

21

13

8

3155

28

#0145

MagPix

48,207

71

1326

1

2

21

13

8

34,823

161

Control

200

49,254

73

1519

1

3

21

13

8

35,446

177

 

FlexMap 3D

49,624

71

1408

2

3

21

13

8

34,831

166

#0150

MagPix

4288

12

204

1

2

21

13

8

8971

12

Control

200

4539

11

204

1

2

21

13

8

8725

12

 

FlexMap 3D

4247

12

204

4

2

21

13

8

8898

12

  1. Shown is output from six representative urine samples, from three confirmed BC subjects (“Tumor”) and three Control subjects. Values are averages of duplicate wells per analyte, per subject, rounded to the nearest whole number. Values that exceeded the highest standard curve concentration for any individual analyte were assigned that biomarker’s highest standard value. The full dataset for all 36 subjects is provided in Additional file 1: Table S1