Skip to main content

Table 4 Precision of reduced hIL-33 iPCR assay

From: Development of an ultra-sensitive human IL-33 biomarker assay for age-related macular degeneration and asthma drug development

Control Mean (pg/mL) %CV
Inter-assaya Intra-assayb
High control 6.52 15 9.4
Mid control 1.82 6.9 13
Low control 0.591 13 8.7
  1. an = 12 runs