From: 9.4 T small animal MRI using clinical components for direct translational studies
Sequence (System) | TR (ms) | TE (ms) | Slice (mm) | Matrix FoV (mm) | BW (Hz/Pixel) | FA (deg.) | Coil | |
---|---|---|---|---|---|---|---|---|
SE (9.4 T Animal) | 1000 | 10 | 0.5 | 128 × 128 64 × 64 | 221 | 90 | Body-boil | Phantom images (Figure 4) |
SE (3 T Human) | 1000 | 10 | 2.5 | 128 × 128 320 × 320 | 221 | 90 | Body-coil | |
GRE (9.4 T Animal) | 95 | 10 | 1 | 256 × 256 50 × 50 | 260 | 25 | Body-coil | Brain images (Figure 5) |
GRE (9.4 T Human) | 95 | 10 | 2 | 256 × 256 250 × 250 | 260 | 25 | 8ch pTX | |
TSE (9.4 T Animal) | 850 | 7 | 0.7 | 384 × 384 60 × 60 | 260 | 140 | Body-coil + 2 channel RX | Spinal cord (Figure 6) |
EPI (9.4 T Animal) | 3000 | 6.2 | 1 | 64 × 64 60 × 60 | 4882 | 35 | Body coil | EPI (Figure 7) |
Multi-GRE (9.4 T Animal) | 75 | 2.5, 5, 7.5, 10, 12,5 | 1 | 64 × 64 60 × 60 | 500 | 7 | Body coil | B0 hom. (Figure 8) |
AFI (9.4 T Animal) | 750 | 2.72 | 1 | 64 × 64 (60 × 60) | 330 | 60 | Body coil | B1 hom. (Figure 8) |