Skip to main content

Table 5 IFNγ real time RT-PCR LOQ and LOD

From: Validation of a HLA-A2 tetramer flow cytometric method, IFNgamma real time RT-PCR, and IFNgamma ELISPOT for detection of immunologic response to gp100 and MelanA/MART-1 in melanoma patients

Expected copies

Detected copies (Mean, n = 12)

SD

Number of positive results/total 12 tests

% Recovery

%CV

IFNγ

  

12/12

  

100,000

104,508

15,676

12/12

105%

15%

10,000

9,032

1,174

12/12

90.3%

13%

1,000

942

198

12/12

94.2%

21%

100

80

27.2

12/12

80%

34%

10

10

NA

8/12

NA

NA

1

1

NA

3/12

NA

NA

CD8

     

100,000

93,334

8,400

12/12

93%

9%

10,000

10,533

2,001

12/12

105%

19%

1,000

1,035

134

12/12

103%

13%

100

109

41.4

12/12

109%

38%

10

14

NA

10/12

NA

NA

1

2

NA

3/12

NA

NA

  1. NA, not applicable (spiked copy number at 10 and 1 did not show detection in all 12 tests, therefore SD, % Recovery, and %CV is not analyzed).
  2. IFNγ plasmid at different copy numbers were used for assessment of LOQ and LOD. Expected and detected values are shown. % Recovery (Detected/Expected) and % CV (SD/Mean) are calculated to assess assay accuracy and precision, respectively.